Lab Scale

Freeze Dryer-FDCF(Chemical-Free Upright Type)

Features

Chemical-free freeze dryer easily captures organic solvents like ethanol, methanol, acetone, hexane, Iso-Octane, etc., which have freezing points of -115℃ ~ -95℃, with the powerful cooling capacity of a cold trap at -120℃ ~ -135℃.
In the case of samples with a freezing point below -100℃, instead of using expensive cryogenic freezers for pre-freezing or liquid nitrogen, self-cooling is possible at temperatures below -120℃ using a stainless 5-tier mini tray (built-in pre-freezer)
Quick, convenient, and very suitable for economically conducting experiments
Minimizes the risk of cross-contamination of samples and situations where the sample does not freeze properly in a general -86℃ freezer, causing it to melt during transfer and fail the experiment
Prevents corrosion caused by solvents by using a Teflon-coated cold trap

-120℃ Chemical-free organic solvent exclusive freeze dryer

Realizes rapid freezing and quick thawing features by applying the Operon Auto Cascade System, a core technology of -156℃ cryogenic refrigeration system and -203℃ cryogenic cooling device.

Technical Specification

Model No. FDCF-12015 FDCF-12008 FDCF-12004
Main Body Cold Trap Temp Below -120℃ Below -120℃ Below -120℃
Capacity (Total) 15L 8L 4L
Dimensions, mm (Exterior) W850 x D850 x H991 W850 x D850 x H991 W800 x D650 x H980
Trap Size, mm Ø340 x L380 Ø310 x L260 Ø310 x L140
Controller 7" TFT LCD Touch Screen, AI (Artificial Intelligence) Advisor, IoT (Internet of Things), USB backup Port (standard 16GB), external output (Ethernet, RS-485, Modbus), analog data external transmission (temperature, vacuum level), real-time ambient temperature detection (monitoring/alarm), real-time voltage detection (monitoring/alarm)<
Pump Protection System Auto/Manual start-up controller, Display cold trap, Temperature & vacumm pressure(2000mTorr~0mTorr)
Defrost Auto / Manual
Safety Device Compressor overload prevention circuit, vacuum pump overload prevention circuit, evaporator abnormality alarm, real-time low voltage alarm, Real-time ambient temperature monitoring and alarm, vacuum leak alarm and control, temperature sensor abnormality alarm, vacuum sensor abnormality alarm
Electrical 220V 1Phase 50/60Hz
Weight 225Kg 216Kg 148Kg

Freeze-drying of ethanol (Test in Operon laboratory)

7-inch Touch Screen

A 7-inch capacitive touch screen is used to implement intuitive control and monitoring.

Trend Graph

Trend view represents the cold trap temperature and vacuum level in a graph, and you can search for past data based on the searched date and time.

Ambient Light & Tilt

The ambient light flashes during an alarm signal. Secures the viewing angle of the touch screen (above 30℃).

Various Interface

USB backup Port (standard 16GB USB provided), external output (Ethernet, RS-485, Modbus), analog data external transmission (temperature, vacuum level)

Internet of Things (IoT)

With the Internet of Things (IoT), you can access your devices from anywhere, anytime using a mobile app. This allows you to control your equipment, monitor its performance, and confirm any alerts through the manufacturer's server.

Optional specifications

Vacuum pump(200LPM)

Drying chamber (acrylic/stainless)

Mini tray4-tier mini tray (pre-freezing possible for cold trap 6,8L)5-tier mini tray (pre-freezing possible for cold trap 12,15L)

T-Manifold, D-Manifold

Cap A,B,C

Flask
300, 700, 900, 1200, 2000ML

Vacuum valve, adapter

Vacuum oil

Vacuum grease

Tube holder

Oil mist trap

Chemical trap

Portable flask stoppering device
(Ø22mm / 10,8,5ml Vials
12holes x 2layers = 24holes)

Chamber stoppering device (Cold trap 4L and above possible)